| IQM 60 |
| Indoor Air Quality Monitor |
| The complete instrument for analysis of indoor air quality. |
| |
The IQM 60 Indoor Air Quality Monitor by Aeroqual enables simultaneous monitoring of common air quality parameters with a single instrument.
Aeroqual's proprietary Analytic Gas Sensitive Semiconductor (GSS) Technology plus photo-ionization detector (PID) and non-dispersive infra-red (NDIR) optical sensors are used to achieve cost effective and precise measurement of Carbon Dioxide (CO2), Carbon Monoxide (CO), Total Volatile Organic Compounds (TVOC), Nitrogen Dioxide (NO2), Ozone (O3), Temperature, Humidity and Particulate Matter.
Air quality data is logged to a removable SD card or PC and can be exported to other programs for generating reports. Applications include IAQ and HVAC analysis, IAQ complaint investigation, and Sick Building Syndrome (SBS) assessment. The IQM 60 offers the complete package for performing in-depth analysis of indoor air quality.
|
 |
| |
| Features |
Applications |
| |
- Multiple Gas Measurement Sensors
- Analytic GSS Technology
- Optical Sensor Technology (PID / NDIR)
- Factory Certified - NIST Traceable
- Temperature and Humidity Sensors
- Particulate Measurement Options - Particulate Profiler (Dual or 8-channel counter) - Particle Monitor (TSP, PM1, PM2.5 or PM10)
- Active-air Sampling System
- Zero and Span Calibration Facility
- Rapid Real-time Data Sampling
- Large Removable Data Storage
- PC Data-logging Software (included)
- Wireless Communication (optional)
- Compact and Portable
- Low Maintenance
|
- IAQ Complaint Investigation and Analysis
- HVAC System Performance Monitoring
- Sick Building Syndrome (SBS) Surveys
- Health and Comfort Assessment
- Odour Investigation and Remediation
- Testing the Efficiency of Air Purifiers
- Clean Room Monitoring, Verification and Filter Testing
- Process Control Monitoring
- Residential and Commercial Buildings
- Airport Lounges and Shopping Malls
- Schools and Kindergartens
- Hospitals and Elderly Care Facilities
- Epidemiological Studies
- Re-entrainment Studies
|
|
|
Gas Sensor Modules (Code)
Carbon dioxide (CD)
Carbon dioxide (CE)
Carbon monoxide (CN)
Total VOC IsoB (VM)
Total VOC IsoB (PD)
Non-methane hydrocarbon (VN)
Hydrogen sulphide (HS)
Nitrogen dioxide (NW)
Ozone (UZ)
Ozone (LZ)
Other calibrations available
|
Sensor
NDIR
NDIR
GSS
GSS
PID
GSS
GSS
GSS
GSS
GSS
|
Sensor Range
0-2000 ppm
0-5000 ppm
0-100 ppm
0-25 ppm
0-20 ppm
0-25 ppm
0-10 ppm
0-0.200 ppm
0-0.150 ppm
0-0.500 ppm
|
Lowest Detection
6 ppm
6 ppm
0.2 ppm
0.1 ppm
10 ppb
0.1 ppm
10 ppb
1 ppb
1 ppb
1 ppb
|
Accuracy
< 40 ppm + 3%
< 150 ppm + 5%
<± 2 ppm <20 <±10% >20 ppm
<± 10%
< 10%
<± 0.5 ppm
<± 0.5 ppm
<± 0.010 ppm
<± 0.005 ppm
<± 0.008 ppm
|
Precision
6 ppm
6 ppm
1 ppm
0.2 ppm
0.02 ppm
0.2 ppm
0.02 ppm
0.005 ppm
0.002 ppm
0.004 ppm
|
Resolution
1 ppm
1 ppm
0.1 ppm
0.1 ppm
0.01 ppm
0.1 ppm
0.01 ppm
0.001 ppm
0.001 ppm
0.001 ppm
|
Temperature and Relative Humidity Probe |
Sensor
Temperature
Relative Humidity |
Range -20°C to 100°C 0-100% RH |
Accuracy ±0.3°C @ 25°C ±2% RH @ 25°C |
Resolution 0.1°C 0.1% RH |
Particulate Profiler Options Right angle light scatter principle
Isokinetic inlet |
Counter Options Dual-channel: >0.3µm |>2.5µm Eight-channel: 0.3µm to 10µm |
Range 0-100,000 particles/L |
Accuracy ±10% to calibration aerosol |
Sensitivity 0.3 µm |
Resolution Particles/L |
Particle Monitor Options Right angle light scatter principle |
Sharp Cut Cyclone Options TSP | PM1 | PM2.5 | PM10 |
Range 0-2000 µg/m3 |
Accuracy 8% of NIOSH 0600 |
Precision 3 µg/m3 |
Resolution 1 µg/m3 |
| Approvals |
EN50082-1:1997 | EN50081-1:1992 Part 15 FCC Rules | IPC A 610D Class 2 |
 |
| Environment Operating Range |
0ºC to 50ºC | 5 to 95% RH (non condensing) |
 |
| Sampling Method |
Diaphragm pump and built-in zero-air scrubber with replaceable media |
| Inlet Filter |
Replaceable PTFE 5 µm |
Zero | Span Calibration |
PC configuration software supplied |
| Communication |
RS 232 | Cable and USB adaptor supplied |
| Data Storage |
Removable SD card 1GB |
| Data Sampling Rate |
Programmable frequency 2-255 minutes |
| Display |
4 x 20 vacuum florescent display (VDF Scrolling data display |
| Power Requirements |
12VDC | 1A (subject to configuration) 100-250 VAC switch-mode PSU supplied |
| Enclosure |
Aluminium and powder coated steel 278L x 236W x 132H (mm) |
| Weight |
3.6 - 7.2 Kg (subject to configuration) |
Communications (optional) |
GSM modem | Wireless |
External Battery (optional) |
Li-ion portable power station and charger (specification subject to configuration)
|
| Accessories |
Carry case | Calibration Kit Calibration Gas Humidifier |
Maintenance Checks Routine |
6-monthly | 12-monthly |
| |